Agilent Technologies 6890 Series Service Manual page 601

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1015
Cabling Diagrams, 6890N
6890N GC GC Automatic Liquid Sampler Non-Agilent Data System
6890N GC
GC Automatic Liquid Sampler
Non-Agilent Data System
Number
1
2
35900-60670 APG remote
cable spade lug identification
Connect 1
9 pin
Signal name
(male)
1
GND
2
Prepare
3
Start
4
Shut down
5
Reserved
6
Power on
7
Ready
8
Stop
9
Start Request
4 of 14
6890N GC
1 AGP remote
Part no. and description
35900-60670, General use APG remote cable, 9-pin male/spade lug (2m)
35900-60920, General use APG remote cable, 9-pin male/spade lug (5m)
35900-60930, General use APG remote cable, 9-pin male/spade lug (0.5m)
G1530-60590, External event cable, 8-pin/spade lugs
Connector 2
spade lugs
Black
White
Red
Green
Brown
Blue
Orange
Yellow
Violet
Cabling Configuration and Electronics
Agilent 6890 Gas Chromatograph Service Manual
The 7683 ALS
controller is internal to
the 6890N GC.
The G2613A injector
and G2614A tray
plug into the GC.
2 External event
Spade lug termination
G1530-60590 External event
cable spade lug identification
Pin
Color
1
Yellow
2
Black
3
Red
4
White
5
Orange
6
Green
7
Brown
8
Blue
Signal
24 V Out 1
24 V Out 2
Ground
Ground
Contact 1
Contact 1
Contact 2
Contact 2
Jun 2001

Advertisement