Figure 430-7
8.
Screw the bracket onto the standoff using a Torx T-10 screw. Do not fully
tighten this screw yet.
9.
From the back of the instrument, secure the bracket with three T-20
screws, starting with the bottom right screw.
Jun 2001
Installing an MIO card (6890A and 6890 Plus)
Installing the MIO support bracket and jumper card
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Power/Electronics Replacement
Insert from rear
of instrument
430
9 of 26