Figure 341-4
Disconnect the µ-ECD vent tube at the rubber sleeve.
6.
7.
Remove the Torx screw holding the gang fitting onto the detector
manifold, and remove the gang fitting.
Remove the one Torx T-20 screw securing the top µ-ECD detector cover
8.
to the detector pallet and remove the cover.
9.
Disconnect the anode signal wire from the detector.
Jun 2001
Micro-cell Electron Capture Detector (µ-ECD)
Disconnecting the make–up gas adapter
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Unscrew Swagelok nut