850
Faults
Fault 232—Back inlet thermal shutdown
Fault 232—Back inlet thermal shutdown
Status message: B inl temp sensor
The back inlet temperature reading reports the inlet's temperature is lower
than expected, which indicates a shorted sensor. Power will be turned off for
inlet.
The sensor must be replaced.
See Fault 221—Thermal shutdown (status message).
Fault 233—Aux 1 thermal shutdown
Status message: Aux l temp too hot
The aux zone's temperature is higher than the maximum allowable
temperature (detector type maximum temperature plus an additional 25° C).
This indicates a defective heater drive on the mainboard or a defective sensor.
Power is disabled to the small zones and the oven. This fault can only be
cleared by turning off all the heated zones and turning the instrument off and
on again.
The problem could be caused by a defective sensor or defective heater
driver on the mainboard.
See Fault 221—Thermal shutdown (status message).
Fault 234—Aux zone 1 thermal shutdown
Status message: Aux l temp sensor
Aux zone 1's temperature reading reports that its temperature is lower than
expected, which indicates a shorted sensor. Power will be turned off for the
zone.
The sensor must be replaced.
See Fault 221—Thermal shutdown (status message).