Oven And Temperature Control Heated Zones; Configuring The Gc For An Msd - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

420
Oven and Temperature Control
Heated zones
the first run of the day. The optional oven insert for fast chromatography
increases oven ramp rates for the back column. Table 420-1 lists typical oven
ramp rates.
Table 420-1
Temperature
range (°C)
50 to 70
70 to 115
115 to 175
175 to 300
300 to 450

Configuring the GC for an MSD

If you are installing an Agilent Mass Selective Detector, you must configure
the GC to properly control the heated transfer line.
1. Press [Config][Aux], and select [1] if the MSD is installed in the front
position or [2] for the back position.
2
Press [Mode/Type].
Use the scroll keys to select MSD as the Aux zone type. Press [Enter].
3
If you do not configure the Aux zone for MSD, Warning 101, Invalid heater
power for front (back) detector, inlet, and aux 1(2), will appear on the GC
display, and the heated zones will be set to not installed.
2 of 20
Oven Ramp Rates
100/120 V oven
ramp rate (°C/minute)
Without
With optional
insert
insert
75
120
45
95
40
65
30
45
20
35
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
200/220/230/240 V oven
ramp rate (°C/minute)
Without
With optional
insert
insert
120
120
95
120
65
110
45
80
35
65
Jun 2001