Agilent Technologies 6890 Series Service Manual page 718

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1230 Power Supply
Setting the instrument power configuration
There are seven possible line voltage power configurations for the 6890 GC.
Table 1230-1
Voltage
Frequency
(–10%, +5%)
(Hz)
120 V
48-66
200 V
48-66
208 V
48-66
220 V
48-66
230 V
48-66
230 V
48-66
240 V
48-66
To change the power configuration for the instrument, you must install the
appropriate types of the following components:
Line voltage configuration plug
Ceramic fuses on the AC power board
Oven shroud assembly
These three components are explained in the following topics.
Jun 2001
Voltage Configuration Information
Maximum power
Power line requirement
consumption (VA)
2,250
20-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,250
10-amp dedicated
receptacle
2,950
16-amp dedicated
receptacle
2,950
13- or 16-amp dedicated
receptacle
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Oven type
Regular
Fast-heating
Fast-heating
Fast-heating
Regular
Fast-heating
Fast-heating
1 of 10

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents