1230 Power Supply
Setting the instrument power configuration
There are seven possible line voltage power configurations for the 6890 GC.
Table 1230-1
Voltage
Frequency
(–10%, +5%)
(Hz)
120 V
48-66
200 V
48-66
208 V
48-66
220 V
48-66
230 V
48-66
230 V
48-66
240 V
48-66
To change the power configuration for the instrument, you must install the
appropriate types of the following components:
•
Line voltage configuration plug
•
Ceramic fuses on the AC power board
•
Oven shroud assembly
These three components are explained in the following topics.
Jun 2001
Voltage Configuration Information
Maximum power
Power line requirement
consumption (VA)
2,250
20-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,950
15-amp dedicated
receptacle
2,250
10-amp dedicated
receptacle
2,950
16-amp dedicated
receptacle
2,950
13- or 16-amp dedicated
receptacle
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Oven type
Regular
Fast-heating
Fast-heating
Fast-heating
Regular
Fast-heating
Fast-heating
1 of 10