Removing The Signal Board - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

340
Electron Capture Detector (ECD)
Replacement procedures

Removing the signal board

1.
Remove both the electronics top cover and the right side cover.
2.
Remove the Torx T-20 screw securing the top cover and remove the cover.
3.
Disconnect the signal wire from the signal board interconnect.
Remove detector
top cover
Remove anode signal wire
Figure 340-11
14 of 24
Removing the top cover and anode signal wire
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001