Shutdowns
840
Shutdown 16—Multiposition valve not switching
The valve could be switching more slowly than usual because it is sticking
slightly or the sample is viscous. Increase the length of the switching time.
7 of 10
Error Codes and Internal Diagnostics
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual