Microcell ECD checkout conditions and chromatogram

Microcell ECD checkout conditions and chromatogram
Column and sample
Type
Sample
Injection volume
Inlet
Temperature
Inlet pressure
Split/Splitless Only
Mode
Purge flow
Purge time
PTV—Splitless Mode
Inlet temperature
Initial time
Rate 1
Final temp 1
Final time 1
Rate 2
Final temp 2
Final time 2
Inlet pressure
Purge time
Purge flow
Detector
Temperature
Const. Makeup Flow, N
Offset
18 of 26
HP-5 30m × 0.32 mm × 0.25 µm
ECD Checkout
1 µL
200°C Purged/Packed
250°C Split/Splitless
Oven TrackCool On-Column
80°C
PTV (see below)
25 psi (Constant pressure for EPC inlets,
helium)
Splitless
60 mL/min
0.75 min
80°C
0.1 min
350°C
2 min
250°C
0 min
0.75 min
60 mL/min
300° C
30 mL/min
2
Should be <400 display units
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
PN 19091J-413
PN 18713-60040
720°C/min
100°C/min
25 psi (const. press. for EPC inlets)