530
Power
Grounding
Table 530-1
Voltage
120 V (±5%)
200 V (±5%)
220 V (±5%)
230 V (±5%)
230 V (±5%)
(Switzerland or Denmark with 10-amp
maximum service)
240 V (±5%)
Frequency range for all voltages is 48 to 66 Hz.
The fast-heating oven requires at least 200 V. Most countries' standard voltage
meets this requirement. GCs for use in the USA, Denmark, and Switzerland
will be equipped with a slow-heating oven unless they are ordered with power
option 002, which specifies a fast-heating oven.
Although your GC should arrive ready for operation in your country, compare
its voltage requirements with those listed in Table 530-2. If the voltage option
you ordered is not suitable for your installation, contact Agilent Technologies.

USA fast heating oven

The fast heating oven requires 240 V/15A power. Do not use 208 V power.
Lower voltage causes slow oven ramps and prevents proper temperature
control. The power cord supplied with your GC is rated for 250 V/15A, and
is a two pole, three wire cord with grounding (type L6-15R/L6-15P).
Receptacle
L6-15R
2 of 6
Line Voltage Requirements
Maximum power
consumption (VA)
2,250
2,950
2,950
2,950
2,250
2,950
Plug
L6-15P
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
Power line
requirement
Oven type
20-amp dedicated
Slow-heating
15-amp dedicated
Fast-heating
15-amp dedicated
Fast-heating
16-amp dedicated
Fast-heating
10-amp dedicated
Slow-heating
13- or 16-amp dedicated
Fast-heating