8.
Loosen the four captive Torx T-20 screws from the top of the detector
pallet and remove the detector from the instrument.
Detector pallet
Figure 310-6
Jun 2001
Removing the detector
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
Replacement procedures
Four captive screws
Ignitor cable
Electrometer ribbon
cable
310
7 of 24