Removing The Fid Interface Board - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

310
Flame Ionization Detector (FID)
Replacement procedures
11. Install the RFI shield, the pneumatics cover, and the detector top cover.
12. Connect the source gas lines to the manifold. See Figure 310-17.
Figure 310-17

Removing the FID interface board

The FID detector has an interface board on the main board as shown on the
following page.
WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off any detector gases at their supply, then turn off the main
power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
1.
Unlock and unclip the ribbon cable from the interface board.
2.
Unclip the FID ignitor from the interface board.
3.
Remove the screw at the top of the card where it mounts to the main
board and pull the card out.
4.
Reassembly is the reverse of removal.
20 of 24
Gas line connections
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001