Removing an EPC flow manifold

Early models of the 6890 GC used Type 1 EPC flow manifolds. Later models
use Type 2 manifolds. Both types are covered here.
The TCD detector EPC flow manifold contains one inlet fitting for reference
and makeup gases. There are three tubes on the pneumatics block fitting going
to the TCD. One is the makeup gas and two are the reference gas.
WARNING
Before proceeding, turn off the detector and any detector gases at their
supply. Then turn off the main power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
1.
Remove the detector cover, the pneumatics cover, metal RFI shield and
the rear top panel.
Jun 2001
Thermal Conductivity Detector (TCD)
Sensor placement in the TCD cell
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures