430
Power/Electronics Replacement
Replacing the ALS Interface Board
5.
Secure the board to the chassis using the two mounting screws. The board
should not be stressed or bowed against the locking tab. See
Figure 430-18.
6.
Locate the 2-wire cable leading from the transformer and connect it to
the ALS Interface board at J5. See Figure 430-18 and Figure 430-16.
7.
Connect the 6890 Controller PCB Cable, part no. G2612-60510, to the
ALS Interface board at P5. If replacing the cable, route the new cable
through the cutout in the main board and connect it to J8. See
Figure 430-16 and Figure 430-17.
8.
If an MIO card was removed from the GC, re-install it.
9.
Re-install the GC covers.
10. Restore power to the GC.
26 of 26
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001