Mounting bracket
Gang fitting
installs here
Label tag through
slot in bracket
Figure 310-13
3.
If the tubing from the gang fitting bends directly to the left, re-shape it
so that it bends up and away from the block as shown in Figure 310-15.
4.
Insert the gang fitting through the cutout in the manifold bracket and
install it onto the new manifold assembly so that the tubing runs back
and away from the fitting. Be sure the left tube clears the inner edge of
the bracket. See Figure 310-13 and Figure 310-15.
Jun 2001
(FPD shown above)
Bracket mounted onto the manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
Replacement procedures
Finger-tighten nuts.
Do not use a wrench.
7/16-inch hex nuts
Ribbon cable
310
17 of 24