Replacing ROMs on the main board (6890A and 6890 Plus)

WARNING
Before proceeding, turn off the main power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
There are four ROM sockets on the lower left corner of the main board
numbered 0 to 3. Only sockets 0 and 1 (unshaded) are currently used.
GND
Lithium
ROM socket 0
battery
ROM socket 1
ROM sockets 2 (top) and 3 (bottom)
Figure 430-9

Removing a ROM

Insert the prongs of an AMP IC puller, part no. 8710-2303
(AMP part no. 821903-1), in the small slots in the upper left and lower right
corners of the socket and pull directly out.
Jun 2001
Replacing ROMs on the main board (6890A and 6890 Plus)
Beeper
P11
P1
P2
GND
P12
Oven Access
Gate
Cutout
Array
P13
CPU
P17
3V
Gate
4 ROM
Array
Sockets
Location of the ROM sockets on the main board
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Power/Electronics Replacement
P21
P16
P22
DSP
P15
F2 F1
+15V
–15V
+5v
Capacitor
430
P3
J1
J2
JP1
JP2
J8
J4
P18
F4 F3
Hole for
J6
transformer
cable
–24V
Capacitors
J5
+24V
11 of 26