Replacing The Als Interface Board - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Replacing the ALS Interface Board

The G2612A ALS Interface board is required to operate the 7683 Automatic
Liquid Sampler. It can only be installed in a 6890 Plus Gas Chromatograph
with a serial number >20,000. While an ALS Interface Board is installed in
the GC, the GC can use only the 7683 ALS.
WARNING
Before proceeding, turn off the main power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
Follow the procedure below to replace the G2612A ALS Interface board.
1.
Remove the GC rear panels and the right side cover.
2.
If an MIO card (LAN card) is installed, loosen its two mounting screws
and slide the card out of the instrument as shown below.
Figure 430-15
Jun 2001
Replacing the ALS Interface Board
Removing the card from the GC
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Power/Electronics Replacement
430
23 of 26

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents