Temperature sensor resistance
The approximate resistance of a temperature sensor is:
R = 100 + (.35 × t)
where R is resistance in ohms and t is sensor temperature in °C.
Table 1240-1
°C
+0°
+10°
100.00
103.90
100°
138.50
142.28
200°
175.84
179.51
300°
212.03
215.58
400°
247.06
250.50
281W
250W
R
E
S
I
200W
S
T
A
N
C
E
150W
(W)
100W
0
Figure 1240-3
Jun 2001
Temperature Sensor Resistance by Heater Temperature
+20°
+30°
+40°
107.79
111.67
115.54
146.06
149.82
153.57
183.17
186.82
190.46
219.13
222.66
226.18
253.34
257.34
260.75
100
°
200
°
TEMPERATURE (
Chart of temperature sensor resistances by heater temperature
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Wiring Harnesses
Temperature sensor resistance
+50°
+60°
+70°
119.40
123.24
127.07
157.32
161.04
164.76
194.08
197.70
201.30
229.69
233.19
236.67
264.14
267.52
270.89
300
°
400
°
°
C
)
1240
+80°
+90°
+100°
130.89
134.70
138.50
168.47
172.16
175.84
204.88
208.46
212.03
240.15
243.61
247.06
274.25
277.60
280.93
500
°