Agilent Technologies 6890 Series Service Manual page 62

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

7.
Note the "Actual" reading on the display and monitor the pressure for
10 minutes.
You can use the stopwatch feature of the 6890 GC to monitor the time.
Press
time and the pressure reading with the
Back Inlet] keys.
[
8.
When the system is considered leak tight, the caps may be removed, the
column reinstalled, its dimensions configured at keyboard, and the
desired pressure and flow rate set.
Jun 2001
Time
and then
Enter
[
]
[
If there is less than 0.5 psi pressure loss (approximately 0.05 psi/
min), consider the system leak tight.
If pressure loss is much greater than 0.5 psi, there is a leak that must
be found and corrected. Note, however, that you may want to slightly
decrease the leak test time based on the internal inlet volume, which
changes with the liner type used (smaller volumes = shorter
acceptable leak test times). See Correcting leaks later in this section.
If there is a rise in pressure, see Forward pressure valve leaks below.
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet
Leak testing—EPC and manual inlets
to start timing, then toggle between the
]
Time
and the
[
]
210
Front Inlet
/
[
]
23 of 30