Agilent Technologies 6890 Series Service Manual page 123

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

240
Programmed Temperature Vaporization Inlet (PTV)
Replacement procedures
Figure 240-7
4.
Remove the column.
5.
Loosen the three captive Torx screws which secure the inlet to the oven,
and lift the PTV inlet out of the oven.
14 of 46
Removing the PTV manifold assembly
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001