Cleaning/Replacing Windows, Filters, And Seals - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

350
Flame Photometric Detector (FPD)
Replacement procedures

Cleaning/replacing windows, filters, and seals

Column bleed and/or effluent can contaminate the first quartz window (heat
shield) nearest the detector module. Dust, fingerprints, atmospheric
contaminants can dirty both quartz windows, the filter, and/or the
photomultiplier tube window. Contamination anywhere along the light path
between flame and PMT can reduce detector sensitivity.
1.
Turn the electrometer off.
2.
Turn hydrogen, air, and auxiliary gas supplies to the detector off. Turn
the heaters off. Wait for the detector to cool.
Caution
Always turn the electrometer off before removing the PMT housing to avoid
destroying the tube.
3.
Release the retaining spring around the photomultiplier housing.
8 of 36
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001