1220 Main Board, 6890A and 6890 Plus
Test points
There are seven test pads on the main board that allow you to test the GC's
regulated DC supplies (+24V, -24V, +15V, -15V, +5V) and to test the ground.
The test pads are located on the main board as shown below:
TP13
GND
Oven Access
Cutout
Figure 1220-1
Jun 2001
TP14
GND
TP
+5V
General specifications
Supply
% tol
+24*
±10%
–24*
±10%
+15
±5%
±5%
–15
±2%
5
±10%
40*
Mainboard test pads
Electrical
Agilent 6890 Gas Chromatograph Service Manual
J8
TP9
+15V
TP10
–15V
Rated I(FL)
System PARD V pp
4.75 A
<1 V
0.75 A
<1 V
0.40 A
<250 mV
0.30 A
<250 mV
4.5 A
<150 mV
11.25 A
(1)
Hole for
Xfmr
cable
TP11
–24V
TP12
+24V
1 of 10