10. Using a pair of needle-nosed pliers, remove the appropriate top rear panel
detector cutout for the µ-ECD. Also remove any cutouts needed to access
other manifolds or accessories installed in the GC. See Figure 341-13.
Front detector
Auxiliary
Left most screw slot
Back detector
Figure 341-13
11. Place the new top rear panel on its left-most mounting screw. Working
from left to right and using the screw as a hinge, slide each manifold ID
tag through its cutout in the panel. When all the tags are through the
panel, finish installing the panel on the GC.
12. Install the RFI shield, the pneumatics cover, and the detector top cover.
Jun 2001
Micro-cell Electron Capture Detector (µ-ECD)
Top rear panel cutouts
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Insert tip of
pliers here
Back inlet
Front inlet