Replacing the active element (bead)
WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off any hydrogen gas at its supply. Then turn off the main power
switch and unplug the power cord.
Caution
To avoid detector contamination, use clean lint-free gloves when handling the
internal detector components. The detector bead is fragile, avoid touching
the bead or letting it touch other surfaces.
While performing this procedure, do the following:
•
Leave all detector gasses, except hydrogen, turned on to purge
contamination. Turn hydrogen off.
•
Make sure the detector has cooled to(<150°C) before beginning, or the
bead may be damaged by exposure to air.
Two ceramic beads are available:
Bead color
White
Black
1.
Turn off the detector active element through the keyboard.
Jun 2001
Nitrogen Phosphorus Detector (NPD)
Part no.
Advantages
G1530-60570
Standard
5183-2007
Durable, no phosphorus
tailing
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Disadvantages
Phosphorus tails
Lower nitrogen
sensitivity,
about 40%
320
13 of 32