Selectivity: 25000 to 1 gN/gC, 75000 to 1 gP/gC with azobenzene/
malathion/octadecane
•
Dynamic range: >10
TCD
•
400° C maximum operating temperature
•
Single filament with microcell
•
MDL: <400 pg propane/mL helium carrier (MDL may be affected by
laboratory environment.)
•
Linear dynamic range: 10
ECD
•
Equipped with anode purge for contamination resistance
•
400° C maximum operating temperature
•
Makeup gas types: argon/5% methane, nitrogen
•
Radioactive source: 15 mCi
•
MDL: <0.04 pg/s lindane
•
Dynamic range: >10
µ-ECD
•
Radioactive source: 15 mCi
FPD
•
Range: 10
•
Selectivity: 10
AED
See the specification guide for AED.
MSD
See the specification guide for MSD.
Jun 2001
5
5
N, >10
P with azobenzene/malathion mixture
5
(± 5%)
63
Ni plated
4
with lindane
63
Ni plated
3
4
(S), 10
(P)
5
6
to 1 gS/gC, 10
to 1 gP/gC
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
Instrument Specifications
Detectors
540
5 of 6