Agilent Technologies 6890 Series Service Manual page 53

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

210
Split/Splitless Inlet
Replacement procedures
Split vent back weldment
Replacement filter kit
Split vent front
weldment
Figure 210-9
WARNING
Turn off the oven and turn off the heater of the inlet you are replacing and
let them cool down. Turn off all detector flows. Turn off the carrier gas supply
pressure, then turn off the main power switch and unplug the power cord.
1.
Disconnect the spit vent trap line from the inlet assembly.
14 of 30
Replacement split vent filter trap kit, part no. G1544-60610
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Filter cartridge
O-rings (2)