Diagnostics; Frequency Test - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Diagnostics

Frequency test

Perform this test to make sure that the base frequency for the ECD during a
blank run indicates a relatively contaminant-free system.
Note
It may take 24 hours for the ECD baseline to completely stabilize, especially
if you are starting with a cold system and want to assure high-sensitivity
operation.
Therefore, for the most accurate results, run the detector at normal operating
conditions for as long as possible (at least 2 hours and up to 24 hours) before
running the frequency test.
If you will be injecting into an unused inlet, you must use low-bleed septa.
Make sure to condition new septa before use in an inlet for several hours with
1 to 5 mL/min carrier flow.
1.
Make sure you are using normal operating conditions and that at least
two hours have elapsed since the last run.
2.
Turn on the ECD and the corresponding signal.
3.
Check the displayed "Output":
Note
Each display count equals a frequency of 5 Hertz (e.g., a display reading of
100 = 500 Hz.).
Jun 2001
<100 = ECD frequency is acceptable
≥100 = Contaminants in system
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Electron Capture Detector (ECD)
Diagnostics
340
17 of 24

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents