Back view of manifold
Figure 330-15
Route tubing along
this path
Figure 330-16
7.
Secure the manifold in place using the Torx T-20 screw.
8.
Route the gas tubing behind the manifold, over the top of the chassis, and
through the slots as shown in Figure 330-16.
Jun 2001
Thermal Conductivity Detector (TCD)
Routing the ribbon cable
Manifold, after installation
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Manifold installed with cable routed to left
Bracket is flush with
carrier rails
Check for interference
Attach gang fitting