Installing A Type 2 Manifold - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

360
Auxiliary EPC
Replacement procedures
Manifold, installed before January 1999
Figure 360-3

Installing a type 2 manifold

If you are replacing a Type 1 manifold with a Type 2 manifold, order the
accessory kit, G1570-60720. This kit contains a new pneumatics control board
bracket that is required for the Type 2 manifold. See also Replacing the PCB
bracket.
WARNING
Hydrogen gas is flammable and potentially explosive. Before replacing the
manifold, turn off the hydrogen gas at the source.
WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off all detector gases at their supply, then turn off the main power
switch and unplug the power cord.
Caution
Always hold the manifold as shown below to avoid damaging board
components.
6 of 16
Remove screw
and gang fitting
Removing the gang fitting
Detectors
Agilent 6890 Gas Chromatograph Service Manual