Removing An Epc Flow Manifold - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

320
Nitrogen Phosphorus Detector (NPD)
Replacement procedures
Lid weldment
Figure 320-10
3.
Loosen the three Torx T-10 screws on the lid weldment and remove the
lid.
4.
Remove one Torx T-20 screw from each end of the electrometer. (You do
not need to remove the screw on the top of the electrometer that holds
the cover on.)
5.
Unlock and detach the electrometer's ribbon cable from the detector's
interface board and lift the electrometer from the detector pallet.
6.
Reassembly is the reverse of removal.

Removing an EPC flow manifold

WARNING
Before proceeding, turn off the detector and any detector gases at their
supply. Then turn off the main power switch and unplug the power cord.
Early models of the 6890 GC used Type 1 EPC flow manifolds. Later models
use the Type 2 manifold. Both types are discussed here.
20 of 32
Removing the NPD electrometer
Detectors
Agilent 6890 Gas Chromatograph Service Manual