Agilent Technologies 6890 Series Service Manual page 719

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1230
Power Supply
Setting the instrument power configuration
Line voltage configuration plug
There is a different line voltage configuration plug on the AC power board for
each power configuration. Each configuration uses three or five jumper wires,
each connecting to two different pins on the plug. The 120 VAC configuration
uses five jumper wires and all other configurations use three jumper wires.
Table 1230-2
Transformer
Oven fan
Ceramic fuses
The two ceramic oven heater fuses on the AC power board (F1, F2) will be
one of two types, depending on the line voltage:
One for the 120 V power option
One for all other power options.
These fuses should always be replaced as a pair.
Power option
120 V
200 V-240 V
2 of 10
Line Voltage Configuration Plugs (P8 on AC board) by Power Option
and Diagram of Pinouts from the Top of the Plug
Jumper locations for power configuration plug
120 VAC
200 VAC
208/220
VAC
3 ↔ 13
2 ↔ 13
3 ↔ 13
6 ↔ 15
6 ↔ 12
6 ↔ 12
9 ↔ 14
1 ↔ 10
1 ↔ 4
1 ↔ 4
4 ↔ 8
Fuse rating
20A/250 V (Type F)
15A/250 V (Type F)
Electrical
Agilent 6890 Gas Chromatograph Service Manual
3
230 VAC
240 VAC
6
5 ↔ 13
3 ↔ 13
9
6 ↔ 9
6 ↔ 9
12
1 ↔ 4
1 ↔ 4
15
2
1
5
4
8
7
11
10
14
13
Jun 2001