Potential Leak Areas - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

250
Volatiles Interface
Leak testing the EPC module only
Disconnect the plumbing at the leaking seal on the interface and examine
the sealing surfaces. Replace the interface if the surface is damaged or
otherwise unable to maintain a good seal. Reassemble and test for leaks.

Potential leak areas

Check the following areas when checking for leaks in the volatiles interface
system.
The capped purge vent
The capped split vent
The plugged column connection
The sealing surfaces where the gas lines are plumbed to the interface
Three O-rings behind the block where the pneumatic lines enter the EPC
module
Two O-rings for each valve in the EPC module
30 of 30
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001