360
Auxiliary EPC
Replacement procedures
15. Attach the Aux EPC label to the top of the pneumatics chassis as shown
in Figure 360-9 below.
Figure 360-9
16. Install the RFI shield, the pneumatics cover, and the detector top cover.
17. If using hydrogen or other flammable gases, install an external flow
restrictor, part no. G1530-80380, on each fitting that will be used for a
flammable gas.
External flow restrictor
installed and ready for
connection to gas line.
18. Check for leaks.
Restore the GC to operating condition
1.
Plug in the GC and turn it on.
12 of 16
Attaching the Aux EPC label
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Attach label here
Gas outlet block
Jun 2001