Split modes
Septum head installed
In the two split modes—with or without a pressure pulse—the solenoid valve
is open and divides the gas stream entering the inlet between the column flow,
the split vent flow through the solenoid valve, and the septum purge flow. The
ratio of the split vent flow to the column flow is called the split ratio.
Figure 240-1 shows the inlet flow control. Overall carrier and sample gas flow
is controlled by the first proportional valve and the flow sensor. The solenoid
valve is open, and proportional valve 2 acts as a back pressure regulator
controlled by the pressure sensor on the septum purge vent line.
Flow
limiting
frit
Figure 240-1
Jun 2001
Programmed Temperature Vaporization Inlet (PTV)
Split
SPR
vent
vent
SPR
PS
Total flow
control loop
Valve
PV2
open
FS
PV1
Temperature
conversion
board
Split mode flow control diagram (septum head installed)
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Septum
holder
Trap
Cryo
valve
240
Split modes
To detector
3 of 46