Mounting bracket
Gang fitting
installs here
Ribbon cable
Label tag through
slot in bracket
Figure 330-13
3.
Peel the blank label from its backing and paste it on the mounting bracket
over the screw heads.
4.
If the original tubing from the gang fitting is not long enough to reach the
new manifold, install the TCD adapter weldment between the gang fitting
and the manifold and shape the tubing so that it bends up and back from
the block as shown in Figure 330-14.
Jun 2001
Thermal Conductivity Detector (TCD)
(FPD shown above)
Bracket mounted onto the manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Finger-tighten nuts.
Do not use a wrench.
7/16-inch hex nuts (3 places)
330
17 of 24