1240
Wiring Harnesses
Inlet/Detector harness (G1530-60640)
Inlet/Detector harness (G1530-60640)
The Inlet/Detector harness runs from connector P21 on the main board to
the front and back inlets and detectors on top of the instrument. It powers
the inlet/detector heater and sensors and the +24 V valve on a manual split/
splitless inlet
.
Wire
color
Black
Green
Yellow
Violet
Gray w/
Violet stripe
Gray
Red
Red w/Orange
stripe
Orange
Blue
White
2 of 8
Pin #
Signal
on P21
name
1
Sensor GND
2
Sensor GND
3
Sensor GND
4
Sensor GND
5
Heater GND
6
Heater GND
7
Heater GND
8
Heater GND
9
+24 Volts
10
+24 Volts
11
Frnt Inj Sense
12
Back Inj Sense
13
Frnt Det Sense
14
Back Det Sense
15
Frnt Inj Htr
16
Back Inj Htr
17
Frnt Det Htr
18
Back Det Htr
19
Frnt Inj +24 Vlv
20
Back Inj +24 Vlv
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Destination
(and Pin #)
FI (3)
BI (3)
FDI (3)
BDI (3)
FI (4)
BI (4)
FD (4)
BD (4)
FV (1)
BV (1)
FI (2)
BI (2)
FD (2)
BD (2)
FI (1)
BI (1)
FD (1)
BD (1)
FV (2)