Agilent Technologies 6890 Series Service Manual page 366

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Route tubing along
this path
Check for interference
Figure 350-17
7.
Route the gas tubing:
8.
Connect the ribbon cable to the mating connector on the pneumatics
board. Arrange the cable to keep it away from the valves and keep it from
being pinched against the manifold.
For the back detector, you may want to loosen the manifold and slide it
out of the carrier a few centimeters to connect the cable to the pneumatics
board. Then reinstall the manifold.
9.
Secure the manifold in place using the Torx T-20 mounting screw.
Jun 2001
Manifold, after installation
If the detector is installed in the back position, route the tubing
behind the manifold, over the top of the chassis, and through the
slots as shown in Figure 350-17.
If the detector is installed in the front position, route the tubing
behind the manifold and over the top of the chassis. Run it directly
to the detector; do not run it through the slots.
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Photometric Detector (FPD)
Replacement procedures
Bracket is flush with
Attach gang fitting
350