Figure 1120-6
10. Place the valve box insulation in the valve box top. Make sure the
perforations in the insulation line up properly with the valve box top.
11. Insert the insulation retainer plate in the bottom of the valve box top and
secure it to the threaded studs using the two 5.5-mm nuts.
Tighten the nuts until they are flush with the top of the stud. Make sure
the holes in the retainer plate line up with the perforations in the
insulation.
Jun 2001
CCW
Stop
Valve in the OFF position (top view)
Valves
Agilent 6890 Gas Chromatograph Service Manual
Valve Box
Installing the valve box
OFF
1120
5 of 10