7.
Unclip the switching valve connector from the detector interface board
(Type 1 manifold) or the in-line connector (Type 2 manifold), and thread
it back up through the mainframe to the pneumatics carrier.
Figure 330-12
Caution
Always hold the Type 2 manifold by its support bracket to avoid damaging
board components.
8.
Slide the manifold from its slot.
Installing a Type 2 manifold
WARNING
Hydrogen gas is flammable and potentially explosive. Before replacing the
manifold, turn off the hydrogen gas at the source.
Jun 2001
Thermal Conductivity Detector (TCD)
Removing the detector interface card
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Switching valve connector
330
15 of 24