Warning 105—Analog out data loss
Status message: Analog out data loss
The 6890 GC analog processors are not communicating with each other
correctly. Analog data will be lost; digital data should be all right.
Although the problem could be caused by a bad power supply or a radio
frequency interference, both are improbable. The most likely cause is a
malfunction on the main PC board.
Warning 106—Non-recoverable data loss. Data corrupt.
Status message: Signal data loss
The 6890 GC digital processors are not communicating with each other
correctly. Both analog and digital data will be lost.
Although the problem could be caused by a bad power supply or a radio
frequency interference, both are improbable. The most likely cause is a
malfunction on the main PC board.
Warning 107—Front det: config changed, method defaulted
Warning 108—Back det: config changed, method defaulted
Status message: F det config changed or B det config changed
This warning appears when the 6890 GC is turned on after a different detector
has been installed, or when a method is loaded that calls for a different type
of detector than the one that is currently installed.
The following will occur:
If a new detector was installed, the 6890 GC will enter the default
setpoints for the new type of detector.
If a method was loaded with a different detector specified, the 6890 GC
will not change the method but will flash the warning message.
Jun 2001
Warning 105—Analog out data loss
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes
830