gang fitting on the flow manifold. Flow control is the same as for the septum
head.
Column head
pressure
control loop
Flow
limiting
frit
Figure 240-3
After the injection, the solenoid valve opens and the flow control is the same
as during split mode operation. See Figure 240-1 (or Figure 240-2).
Jun 2001
Programmed Temperature Vaporization Inlet (PTV)
Split
SPR
vent
vent
SPR
PS
Valve
PV2
closed
FS
PV1
Temperature
conversion
board
valve
Splitless mode flow control diagram: sample injection
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Septum
holder
Trap
Cryo
240
Splitless modes
To detector
7 of 46