Split/Splitless Inlet
Leak testing—EPC and manual inlets
Figure 210-14
Note
Make sure that the carrier gas source pressure is at least 35 psi. Carrier source
pressure should always be at least 10 psi greater than the desired inlet
pressure.
20 of 30
Manual
EPC
Capping the bottom of the inlet and septum purge vent
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Septum purge vent
Septum purge vent
Split vent