6890N
The connector provides the control relays and a BCD input for a stream
selector Multi Valve.
8
2
BCD Connector
Pin
1
2
3
4
5
6
7
8
Shield
Jun 2001
6
1
Function
Color
Relay
Yellow
Relay
Black
LS digit 0
Red
LS digit 1
White
LS digit 2
Orange
LS digit 3
Green
MS digit 0
Brown or Violet
GND
Blue
Chassis GND
Cabling Configuration and Electronics
Agilent 6890 Gas Chromatograph Service Manual
Cable Electronics
BCD Inputs
Maximum rating
48 V AC/DC, 250 mA
48 V AC/DC, 250 mA
1020
7 of 18