Mounting bracket
Gang fitting installs here
Ribbon cable
Figure 350-15
3.
Peel the blank label from its backing and paste it on the mounting bracket
over the screw heads. See Figure 350-16.
4.
If the tubing from the gang fitting bends to the left, reshape it so that it
bends up and back from the block as shown in Figure 350-17.
5.
Insert the gang fitting through the cutout in the manifold bracket and
install it onto the new manifold assembly so that the tubing runs back
and away from the fitting.
Jun 2001
Bracket mounted onto the manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Photometric Detector (FPD)
Replacement procedures
Finger-tighten nuts.
Do not use a wrench.
7/16-inch hex nuts
Label ID tag through
slot in bracket
350
27 of 36