Split/Splitless Inlet
210
Replacement procedures
10. Reinstall the blue inlet plastic cover or tray bracket to the top of the inlet
area.
OR
Figure 210-8
Reinstalling the inlet cover or tray bracket

Replacing the split vent trap

For GCs manufactured before May 1997, the split vent trap (part no. G1544-
80550) or EPC units can be replaced by the Replacement Split Vent Trap kit,
part no. G1544-60610. The new split vent trap uses a replaceable filter
cartridge, which is sold in a package of 2 each (part no. G1544-80530).
To replace the filter cartridge in the trap, see Replacing the split vent trap
filter cartridge below.