Figure 430-4
d.
8.
Remove the three 1/4-inch hex screws and the two Torx T-20 grounding
screws holding the board to the instrument body.
Jun 2001
Removing the MIO support bracket and jumper card
Pull out the MIO jumper card (P15) from the main board.
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Power/Electronics Replacement
Replacing the main board
Jumper card
430
Remove
Standoff
5 of 26