Replacing The Split Vent Trap Filter Cartridge - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

EPC pneumatics block
HPM8 fitting
Old trap assembly
1/8-inch OD fitting
Figure 210-10
2.
For EPC inlets, follow the split vent line from the trap to the HPM8
connector from the pneumatics block on the EPC manifold. Disconnect
the plastic fitting and remove the old trap assembly and connected
plumbing.

Replacing the split vent trap filter cartridge

1.
Turn off the inlet and the oven and allow to cool.
2.
Set all GC flows to zero.
3.
Remove the pneumatics cover.
4.
Lift the filter trap assembly form the mounting bracket and unscrew the
filter trap assembly.
5.
Remove the old filter cartridge and O-rings and replace them.
Jun 2001
EPC manifold configuration
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet
Replacement procedures
210
15 of 30