430
Power/Electronics Replacement
Replacing the transformer—GC serial number ≥ 10225
3.
Unclip the transformer wiring harness from connector J1 on the AC
power board.
4.
Unclip the connector from P9 on the AC power board and move the wires
out of the way of the transformer.
20 of 26
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001