Agilent Technologies 6890 Series Service Manual page 274

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1.
Remove the detector carrier bracket and detector cell as described in the
Replacing the entire detector procedure.
2.
Remove the cover and the insulation.
3.
Use a T-20 Torx screwdriver to remove the two screws securing the
detector cell to the aluminum detector carrier bracket and lift the cell
from the bracket.
Figure 330-9
4.
Slide the heater/sensor assembly from the detector cell.
If your TCD uses a ∆
5.
Jun 2001
Thermal Conductivity Detector (TCD)
Detector cover
Insulation
Heater/sensor assembly
Detector cell
Insulation
Aluminum detector bracket
Insulation
Removing the detector cell from the carrier bracket
PRT, slide it out of the detector cell.
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
D PRT
330
11 of 24