Replacing The Supply Fitting On A Type 1 Epc Manifold - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

5.
Remove the plumbing block from the front of the manifold (one Torx T-10
screw).
6.
Remove the long screw (Torx T-20) from the top of the manifold and slide
the manifold out of the back of the GC.
7.
Reinstallation is the reverse of removal.

Replacing the supply fitting on a Type 1 EPC manifold

The same replacement procedure is used to replace the supply fitting on any
of the inlet flow inlet manifolds used in the 6890 GC. See Split/Splitless Inlet
for the detailed procedure.
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed Inlet
Replacement procedures
220
9 of 20

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents